  1. MPC-102S

    Model: MPC-102S
    Test Standards: PP: ASTM D6749, ASTM D97, ISO 3016; CP: ASTM D7683, ASTM D2500, ISO 3015
    Specimen Volume: 4.5ml
    Measurement Modes: Selectable from various modes. 1.CP mode (0.1 or 1.0 °C intervals) 2.PP mode: Programmed by the user. Programmable parameters are: *Amount of applied air pressure for PP detection, to accommodate different sample types: L(low) for diesel fuels, H(high) for lube oils, VH(very high) and UH (ultra high) for residual fuels and similar samples. *Testing intervals: 1.0°C, 2.5°C, or 3.0°C (In total, 4x3=12 modes for PP.) 3.CP/PP modes: CP is determined and then PP. PP detection is programmable by the user with the same parameters as PP modes.
    Measuring Range: -65°C to +51°C
    Sample Automatic Pre-Heating: Automatic preheating at either +45°C or EPP+10°C. (EPP=Expected Pour Point)
    Display: Big digital display on VFD for showing real time parameters (sample temperature, bath temperature, test results).
    EPP Setting: EPP (Expected Pour Point) needs to be set prior to test starts.
    Specimen Cup: Cylindrical glass test jar with 4.5ml sample volume with reflex seal on the bottom for cloud point test.
    Sensors: Compound type sensor assembly for PP and CP. PP detected by air pressure method (patented). CP detected photo-electrically. PT100 temp. sensors.
    Sample Cooling Rate: As standard, 4°C/min. till EPP+40°C, and 1°C/min. Thereafter cooling profile is programmable.
    Safety Shutdown: As hot side of TED reaches 60°C while preheating, warning buzzer beeps and heating stops.
    Data Output: RS-232C 1 channel (for PC or Optional Printer)
    Power Supply: 100VAC to 240VAC 50/60Hz
    Power Consumption: 80 Wh for 1 test (about 1 hour), 30g CO2 (@0,378 Kg CO2 / 1000 Wh)
    Operating Ambiental Temperature: 10 to 35°C
    Size (W x D x H): 300 x 460 x 320 (mm)
    Weight: 11 Kg

    Bath Model: CB-80C
    Bath Temperature Range: -70°C to +80°C
    Bath Temperature Control: PID control with 2 points of revision functions
    Bath Size (W x D x H): 222 x 407 x 238 (mm)
    Bath Weight: 13 Kg

    IQ/OQ/PQ Qualification: Yes, possible
  2. MPC-102L

    Model: MPC-102L
    Type:: Mini Pour Point(PP) and Cloud Point(CP) tester with sequential CP and PP measuring capability. Sample cooling and pre-heating by TED.
    Test Standards: ASTM D6749/D97, ISO 3016 (PP), ASTM D7683/D2500, ISO 3015 (CP)
    Specimen Volume: 4.5ml
    Cooling: Liquid Cooling
    Measuring Range: -40°C to +51°C with tap water of 20°C; -65°C to +51°C with cooling liquid of -35°C (*:Sample viscosity, etc. affects on lowest temperature of the measuring range.)
    Measurement Modes: Selectable from various modes. 1.CP mode (0.1 or 1.0 °C, selectable) 2.PP modes: Programmed by the user. Programmable parameters are: *Amount of applied air pressure for PP detection, to accommodate different sample types: L(low) for diesel fuels, H(high) for lube oils, VH(very high) and UH (ultra high) for residual fuels and similar samples. *Testing intervals: 1.0°C, 2.5°C, or 3.0°C (In total, 4x3=12 modes for PP.) 3.CP/PP modes: CP is determined and then PP. PP detection is programmable by the user with the same parameters as PP modes'. (12 modes in total.)
    Sample Automatic Pre-Heating: Automatic preheating at either +45°C or EPP+10°C. (EPP=Expected Pour Point)
    Display: Test parameters, EPP, bath temperature, sample temperature, PP, and CP displayed on VFD. Temperatures displayed in 0.1°C increments.
    EPP Setting: EPP (Expected Pour Point) needs to be set prior to test starts.
    Specimen Cup: Cylindrical glass test jar with 4.5ml sample volume.
    Sensors: Compound type sensor assembly for PP and CP. PP detected by air pressure method (patented). CP detected photo-electrically. Pt100 temperature sensors.
    Sample Cooling Rate: As standard, 4°C/min. till EPP+40°C, and 1°C/min. Thereafter, cooling profile is programmable.
    Safety Shutdown: As hot side of TED reaches 60°C while preheating, warning buzzer beeps and heating stops.
    Data Output: RS-232C 1 channel (for PC or Optional Printer)
    Data Strage: Last 50 test data are stored in RAM
    Power Requirements: 100, 120, 220, or 240VAC 0.5kW
    Dimensions (WxDxH) (mm) and Weight: 230mm x 480 x 385, 20kg

    IQ/OQ/PQ Qualification: Yes, possible
  3. MPC-102A

    Model: MPC-102A
    Type:: Mini Pour Point(PP) and Cloud Point(CP) tester with sequential CP and PP measuring capability. Sample cooling and pre-heating by TED.
    Test Standards: ASTM D6749/D97, ISO 3016 (PP), ASTM D7683/D2500, ISO 3015 (CP)
    Specimen Volume: 4.5ml
    Cooling: Air cooling
    Measuring Range: -30°C to +51°C (in 25°C ambient) (*:Sample viscosity, etc. affects on lowest temperature of the measuring range.)
    Measurement Modes: Selectable from various modes. 1.CP mode (0.1 or 1.0 °C, selectable) 2.PP modes: Programmed by the user. Programmable parameters are: *Amount of applied air pressure for PP detection, to accommodate different sample types: L(low) for diesel fuels, H(high) for lube oils, VH(very high) and UH (ultra high) for residual fuels and similar samples. *Testing intervals: 1.0°C, 2.5°C, or 3.0°C (In total, 4x3=12 modes for PP.) 3.CP/PP modes: CP is determined and then PP. PP detection is programmable by the user with the same parameters as PP modes'. (12 modes in total.)
    Sample Automatic Pre-Heating: Automatic preheating at either +45°C or EPP+10°C. (EPP=Expected Pour Point)
    Display: Test parameters, EPP, bath temperature, sample temperature, PP, and CP displayed on VFD. Temperatures displayed in 0.1°C increments.
    EPP Setting: EPP (Expected Pour Point) needs to be set prior to test starts.
    Specimen Cup: Cylindrical glass test jar with 4.5ml sample volume.
    Sensors: Compound type sensor assembly for PP and CP. PP detected by air pressure method (patented). CP detected photo-electrically. Pt100 temperature sensors.
    Sample Cooling Rate: As standard, 4°C/min. till EPP+40°C, and 1°C/min. Thereafter, cooling profile is programmable.
    Safety Shutdown: As hot side of TED reaches 60°C while preheating, warning buzzer beeps and heating stops.
    Data Output: RS-232C 1 channel (for PC or Optional Printer)
    Data Strage: Last 50 test data are stored in RAM
    Power Requirements: 100, 120, 220, or 240VAC 0.5kW
    Dimensions (WxDxH) (mm) and Weight: 230mm x 480 x 385, 20kg

    IQ/OQ/PQ Qualification: Yes, possible
  4. ASP-6

    Model: AAP-6
    Test Methods: ASTM D36, ASTM E28, EN 1427, IP 58, etc
    Type: Two-rings & two-balls
    Measuring Range: Ambient to 200?C (392?F )
    Temperature Sensor: PT-100 in stainless steel sheath
    Softening Point Detector: Light projector: LED; Light receiver: photo-diode
    Display: 5.7" color LCD with back light
    Heater: Nichrome coil heater: 1,000W @100V or 1,000W @220V
    Sequense Control: By an on-board microcomputer
    Test Mode: ASTM D36 water, ASTM D36 glycerin, EN 1427 water, EN 1427 glycerin
    Stirring: Magnetic stirring, selectable from 0 to 250 rpm
    I/O Port: 1x Ethernet for LIMS; 1x USB for Flash Memory or Keyboard; 1x RS-232C for LIMS / PC or Optional Printer
    Data Memory: Last 200 sets of test results
    Safety Mechanism: a) The sample falls on only one side and not on the other within the set temperature range, b) The sample does not fall on either side within the set upper limit temperature, c) Light projector or receiver is found defective, d) Temperature sensor is found defective, e) Temperature sensor is not placed in the right position, f) Built-in battery is found drained out, or g)Control computer runs away (no display)
    Size (WxDxH): 230 x 470 x 440 (mm)
    Weight: 15 kg
    Ambiental Conditions: RT: 10 to 40?C, RH: Less than 90% (no condensation)
    Power Supply: AC100/120 V or 220/240V 50/60Hz (set at the factory, specify when ordering) 1,500W ( max.)

    IQ/OQ/PQ Qualification: Yes, possible
  5. AAP-6

    Model: AAP-6
    Conforming Standards: ISO2977, ASTM D611, etc.
    Measuring Range: Ambient +20°C to 170°C, for bellow ambient temperature + 20°C, the optional water jacketed measuring cell and a cryostat must be used.
    Temeprature Sensor: Pt100 in stainless steel sheath
    Aniline Point Detector: Photo-electric cell with special detection circuitry
    Measuring Cell: Modified U-Tube type, vacuum jacketed glass cell with PTFE stopcock at the bottom for drainage.
    Display: 5,7" color LCD with LED backlight
    Heater: Immersion heater in stainless steel sheath: 20W
    Sequense Control: By an on-board microcomputer
    Test Mode: Sinle, repeat (2 kinds, average is reported to nearest 0,05°C), SPE and manual temp set
    I/O Port: 1x Ethernet for LIMS, 1x USB for Flash memory or keyboard, 1x RS-232C for PC or optional printer
    Safety Mechanism: Automatically shuts off and the problem is reported by buzzer and display in case: a) EAP+10°C or 170°C is reached, b) Temperature sensor is found defective, c) Light projector or receiver is found defective, d) Built in battery is found drained out, e) Control computer runs away (no display)
    Dimensions (W X D X H): 230 x 470 x 490 mm
    Weight: 14Kg
    Ambiental Conditions: 10 to 40°C, Relative Humidity less than 90%
    Power Supply: AC 100 to 240V; 50/60Hz; 125W max

    IQ/OQ/PQ Qualification: Yes, possible
  6. AKV-202

    Model: AKV-202
    Type: Bench top automatic KV measuring system with 2 viscometers in 1 bath
    Related Standards: ISO 3104, ASTM D445, IP71, etc.
    Viscometers: Modified Lanz-Zeitfuchs viscometers
    Measuring Range: 10,000 mm2 /s for automatic operation. (Range depends on viscometer.)
    Meniscus Detection: By photo-electric devices with fiber optics
    Timing: 0.00 - 999.99 sec
    Bath: Stainless Steel bath with drain, 18 liters
    Bath Heater: Sheathed heaters, 0.59kW x2
    Bath Lighting: Fluorescent light, 6W
    Bath Medium: Water (distilled or ion exchanged) for < 75°C, Colorless/transparent silicon oil (with PMCC flash point of 190°C at minimum and KV of 10-20mm2/s) for >= 75°C.
    Bath Temperature Controller: Temp: Preset at 6 points. Preset point can be set at any temp between 20 and 100°C; Precision: ±0.01°C
    Cleaning/Drying Viscometers: An automatic cycle consisting of draining sample, injecting chemicals, and drying with air. By vacuum; Chemicals: BTX for cleaning, Ethanol ?Bath Temp <80°C?for drying. (Ether or Ketone can not be used); Fume: Fume to be exhausted through trap and hose to outside
    Display: Test parameters, status, trouble, etc. displayed on fluorescence display module
    Printer: Thermal printer, 58mm of paper width
    Computer Output: RS-232C, 1 channel
    Safety: Alarms for gas, bath medium level, & bath overheat
    Power Requirements: 100, 120, 220 or 240VAC, 2 KW
    Dimension & Weight (Wxdxh): 530mm x 560mm x 930mm, 75kg

    IQ/OQ/PQ Qualification: Yes, possible
  7. AVP-30D4

    Model: AVP-30D4
    Related Standards: ISO 3007, ASTM D323, IP69 etc.
    Type: Bench-top automated instrument. Automated continuous shaking of test cylinder by "seesaw" motion.
    Number Of Cylinders (Measurement Spots): 4
    Measuring Range: 0 to 196 kPa (2 kgf/cm2)
    Test Cylinders: Small size stainless steel cylinders with 66% in length and 31% in volume compared to the regular full size cylinders. The V/L ratio of 4 is the same as full size cylinder. A "Quick Coupler" is equipped at the center of the vapor cylinder for connecting to external pressure sensor located inside of the tester
    Pressure Sensor: Semiconductor type pressure sensor.
    Range: 0 to 196 kPa (2kgf/cm2)
    Linearity: <0.3% of the range
    Hysteresis: <0.3% of the range
    Display: LCD with back light
    Pressure Display Range: 0 to 200 kPa or 0 to 2 kgf/cm2
    Shaking: By "see-saw" movement with angular movement of +/-15 degrees and interval of 6 sec
    Bath Temperature Range: Ambient to 60°C
    Temp Control: Digital controller/display with PT-100 Sensor, Precision: +/-0.1°C
    Stirring: By propeller driven by motor
    Heater: Stainless sheathed heater 1.15kW
    Calibration: By connecting a Digital Manometer (optional) to the pressure sensor
    Safety Mechanism: Water level monitoring device for preventing overheating due to low water level
    Data Output: 1xRS-232C channel (for PC or Optional Printer)
    Dimension & Weight: 400W x 600D x 720H (mm), 50kg when dry, 85kg when filled with water
    Installation Site: Ambient temperature: 0 to 40°C
    Relative Humidity: Less than 90% (no condensation)
    Power Supply: AC100/120V or 220/240V 50/60Hz (set at the factory) 1.5 kW (max.)

    IQ/OQ/PQ Qualification: Yes, possible
  8. TCU-40B

    Model: TCU-40B
    Temperature Range: -40°C to Ambient
    Temperature Control: By digital display controller, ON/OFF control
    Temperature Stability: +/- 1°C
    Cooling Capacity: 87W@ -40°C, 158W@ -30°C, 212W@ -20°C (@50Hz)
    Pumping Capacity: Max Head: 2.0m, Max Flow Rate: 10lit/min, Force type
    Bath Volume: 7 litre
    EPP Setting: Current Leakage Breaker
    Power Source Voltage: AC 120V, 220V-240V *Specify when ordering
    Power Consumption: 700W max.
    External Dimensions: 320(W) x 590(D) x 725(H) mm
    Weight: 55kg (AC120V); 60kg (AC220-240V)
  9. AFP-102

    Model: AFP-102
    Conforming Standards: IP309, ASTM D6371, EN116 and JIS K2288
    Type: 1 test head.
    Specimen Volume: 45ml
    Measuring Range (Typical): RT to -60°C (The lowest temperature varies with the chiller liquid temperature and ambient temperature.)
    Sequence Control: By a built-in micro processor
    Display: By a fluorescence display (VFD) Temperatures displayed with 0.1°C increments
    Temperature Sensor: Pt-100
    Cfpp Detection: By means of a photo-electric detector consisting of an LED, photo transistor, and light guide (fiber optics). Upper and lower light guides are mounted on sliding type holder.
    Vacuum Regulator: Consists of a micro air pump, pressure switch, solenoid valve and a plastic bottle.
    Cooling And Heating (Needs Optional Chiller): By Peltier cells. The hot side of the Peltier cells needs to be cooled by external chiller.
    Temperature Control: By step (-34°C, -51°C and -67°C) or linear. The linear cooling rate can be programmed.
    Safety Features: a) Over cut: as hot side of TED reaches 60°C, warning buzzer beeps and stops. b) Self diagnostic: reported by an intermittent buzzer and a text message when any temperature sensor breaks down or the battery has drained out. c) System trouble: reported by a buzzer when the built-in microprocessor runs off.
    Data Output: 1x RS-232C (for PC or Optional Printer). 2nd set channel is optional.
    Power Requirements: 100 or 120VAC, 5A; 220 or 240VAC, 3A Please specify AC voltage when ordering.
    Dimensions And Weight: 350W x 550D x 480H mm, 27kg

    IQ/OQ/PQ Qualification: Yes, possible
  10. PX224

    Model: PX224
    Kataloški broj: 30429804
    Odobren model (kod Direkcije za mere i dragocene metale): Da
    Kalibracija: Interna
    Kapacitet (g): 220
    Rezolucija (g): 0,0001
    Ponovljivost (STDEV) (g): 0,0001
    Linearnost (g): ±0,0002
    Vreme stabilizacije (s): 4
    Osetljivost temperaturnog odstupanja (PPM/K): ± 3
    Tipična minimalna masa merenja prema USP (USP K=2, U=0,10%): 200 mg
    Optimizovana minimalna masa merenja (USP, U=0,10%, K=2) SRP≤0,41d*: 82 mg
    Jedinice merenja: Milligram, Gram, Kilogram, Ounce, Pound, Carat, Pennyweight, Ounce Troy, Grain, Newton, Hong Kong Tael, Singapore Tael, Taiwan Tael, Momme, Tical (MM), Mesghal, Tola (India), 1 Custom unit
    Aplikacije: Osnovno merenje mase, Brojanje komada, Procentulano merenje, Merenje životinja, Određivanje gustine
    Dimenzije merene ploče (Ø): 90 mm
    Napajanje: Ulaz: 100-240V; 200mA 50-60Hz 12-18VA, Izlaz: 12VDC 0,5A
    Dimenzije vage (Š x D x V): 209 x 321 x 309 mm
    Komunikacioni interfejsi: RS232 i USB
    Opseg radne temperature: 10°C do 30°C
    Neto težina: 4,5 kg
    Težina transportnog pakovanja: 7 kg
    Dimenzije transportnog pakovanja (Š x D x V): 507 x 387 x 531 mm

    IQ/OQ/PQ Kvalifikacija: Da, moguća je

    *: Vrednost SRP je standardna devijacija za n broj merenja (n≥10)